Portable Sampling Probe
Features
☆ Hidden quick joint, easy installation, and good heat preservation effect
☆ High quality filter element from Germany
☆ Double filtration, more safe to the analyzer
Model |
PW-SP-P series |
Max gas flow |
<4LPM (Max humidity 40V/V) |
Pre-heating time |
<20min |
Max gas tempreture for inlet |
200°C ( other temperatures optional) |
Probe connetion |
Hidden quick joint |
Outlet connection of sample gas |
6mm PTFE pipe |
Electrical requirment |
220V AC 550W |
Material for the pipe |
316 SS & PTFE |
Size |
Probing rod: 1.5m (1.2m optional) |
Flexible heating line: 3m |
|
Filter element |
Front filter element: 5um |
Inside filter element: 0.lum |
|
Heating temperature |
180°C |
Weight |
< 5kgs ( net weight) |
Working environment |
-20~50℃ |
Principle
After the primary filter on the probe rod, the flue gas enters the sampling chamber for secondary precision filtration, then enters the heat tracing pipeline through the sampling stop valve, and finally is sent to the analyzer